Ⅰ. Introduction
High-resolution magnetic resonance imaging(MRI) is a typical method for examining the human body by using nonionizing radiation. High-resolution MRI is a method using a magnetic field and radio frequency(RF) that are harmless to the human body, which is characterized by excellent spatial resolution and contrast resolution[1, 2]. The space in which the MRI stores data is called k-space. Among the various methods to obtain the data by filling the data in the k-space, the cartesian method, which is a scanning method in the cartesian coordinate system, is to fill the data in the frequency encoding direction[3]. The advantage of cartesian sampling is that the contrast of the image can be easily adjusted through the effective time of echo(TE) of the array corresponding to the contrast of the k-space[4-7]. However, as each row is linearly acquired in proportion to the number of phase encoding arrays, the involuntary motions of patients lead to various artifacts in the phase encoding direction[8]. As a result, artifacts are generated linearly throughout the image, so that the target region to be viewed is covered with artifacts hindering accurate images from being obtained. The MultiVane scanning method in a nonrectangular coordinate system is a noncartesian technique that oversamples the center of the k-space using separate vanes[9-12]. Each vane includes a k-space sample corresponding to the same low-resolution image. The data are collected in a series of revolving vanes, and each revolving vane collects the data from the central region of the k-space[13, 14]. The MRI phantom and protocol of the American College of Radiology(ACR) were developed by the ACR in 1992, and this MRI phantom could be applied to all types of MRI equipment[15]. Quantitative tests such as geometric accuracy, high-contrast spatial resolution, image intensity uniformity, and low-contrast object detectability have been performed by using ACR phantom[16]. In actual clinical setting, the MultiVane scanning method is used for motion compensation. This study was performed a comparative analysis of differences between cartesian trajectory and MultiVane trajectory in axial T1 and axial T2 images by using the ACR phantom to provide the basic clinical data.
Ⅱ. Material and Methods
1. Material and equipment
The equipment used in this study includes a 3.0 T MRI(Achieva, Philips, Netherlands) and the ACR phantom. The ACR phantom is an acrylic plastic with both ends closed, having a length of 148 mm and a diameter of 190 mm. The inside of the phantom is filled with a solution of 10 mM of nickel chloride (NiCl2) and 75 mM of sodium chloride (NaCl). The ACR phantom can be divided into a large phantom and a small phantom. The large phantom is designed for use in a head coil, which is used for quality control of most MRI equipment. The small phantom is designed for use in a knee coil[17]. As shown in Figure 1, the surface of the large phantom is marked with NOSE and CHIN, which helps determine the location of the phantom in the head coil[16, 18].
As show in Table 1, the used parameters were the same as those used in the actual clinical setting.
2. Measurement Methods
The phantom was placed at the center of the head coil and the top-to-bottom and left-to-right levels were adjusted by using a level. The experiment was performed according to the Phantom Test Guidance provided by the ACR, and sagittal localizer images were obtained. As shown in Figure 2, slices # 1 and # 11 were scanned after placing them at the center of a 45° wedge shape, and a total of 11 slices were obtained. Multivane trajectory and Cartesian trajectory were compared in actual brain axial T1 images.
The geometric accuracy is to evaluate whether the length in the image represents the actual length of the object. The lengths of slices # 1 and # 5 were measured in axial T1 and T2 in left-to-right, top-to-bottom, and both diagonal directions. The normal range was set as 190±2 mm for the inner diameter of the phantom. The high-contrast spatial resolution is to evaluate the ability to analyze small objects when the contrast-to-noise ratio(CNR) is sufficiently high. The resolution insert consists of an upper left(UL) and a lower right(LR) with diameters of 1.1 mm on the left, 1.0 mm in the middle, and 0.9 mm on the right. The normal range in this assessment is ≤ 1.0 mm. The arrangement of the resolution insert is composed of upper left(UL) and lower right(LR). The result values of image intensity uniformity are represented as the value calculated according to the percent integral uniformity(PIU) formula[16] as shown in Equation(1).
A region of interest (ROI) of 190–205 cm² was draw in slice #7. The window width and window level was adjusted to display the highest part and the lowest part of the signal intensity (mean pixel value) as an area of interest of 1 cm² in the whole area of interest. The value obtained by substituting the measured value at that part into the PIU formula corresponds to the IIU. The normal range is equal to or greater than 82%. Low-contrast object detectability is used to assess by what degree of contrast an object can be distinguished in an image. In the case of 3 T, the normal range for the low-contrast object detectability is 37 spokes.
3. Motion compensation in MultiVane
The motion compensation algorithm used in MultiVane consists of several imaging correction steps. First, the average blade(RB0) of all the blades that were not corrected is calculated. Second, a single blade(RB1) showing the best correlation with RB0 is selected, a correlation with RB1 is measured for each of the other blades, and the expected shift is applied to each blade. This average is formed after imparting the first weight to each blade according to the correlation with RB1. Third, all the blades including RB1 are relocated, and a new average blade(RB2) is repeatedly generated. This process is iterated for the average blade RBi until the last motion compensation is completed for the average blade RBi+1. As shown in Figure 3, N blades and k-space with a diameter of L are resampled as the blade for every strip rotating with respect to the center of the k-space, and the data are acquired by superimposing with the strip. Assuming that the two blades are fi and fj and that the streaks from the X axis δi and δj the correlation between these two blades can be calculated based on Perceval's theorem, which is shown in the Equation (2).
where Fi and Fj are the results of the Fourier transforms for fi and fj respectively, and the relative motion of (δi-δj)can be predicted by the maximum correlation between the two blades[20].
Ⅲ. Results
Table 2 shows the results of geometric accuracy. The length values for both cartesian and MultiVane trajectories are within the normal range of 190±2 mm. Table 3 shows the results of high-contrast spatial resolution. The values in the cartesian trajectory were measured as 0.9 mm in both axial T1 and T2. The results were within the normal range. On the other hand, the values in MultiVane were 1.0 mm in axial T1, 0.9 mm for UL. However that of MultiVane was 1.1 mm for RL in axial T2 which was deviated from the normal range.
Figure 4 shows the results for the IIU. The normal range is equal to or greater than 82%. In axial T1, the IIU in the cartesian trajectory is 93.34%, and the IIU in the MultiVane trajectory is 93.19%, which all fall under the normal range. However, the IIU in the cartesian trajectory is by 0.15% higher than that of the MultiVane trajectory. In axial T2, the IIU in the cartesian trajectory is 96.44% and the IIU in the MultiVane trajectory is 95.97%, which all fall under the normal range. However, the IIU in the cartesian trajectory is by 0.47% higher than that of the MultiVane trajectory.
Figure 5 shows low-contrast object detectability. In axial T1, the value in the cartesian trajectory was 38.4 spokes, which falls under the normal range, and the value in the MultiVane trajectory was 36.2 spokes, which deviated from the normal range. In axial T2, the value in the cartesian trajectory was 37.2 spokes cartesian, and the value in the MultiVane trajectory was 31.2 spokes. Thus, the value in the cartesian trajectory falls under the normal range, while the value in the MultiVane trajectory was deviated from the normal range. Furthermore, the value in the MultiVane trajectory was deviated from the normal range by 2.16% in axial T1, and by 15.67% in the axial T2.
Figure 6 shows that cartesian images exhibit better contrast and sharpness than MultiVane images in actual brain MRI axial T1 images. Furthermore, the cartesian images are more clearly distinguished between gray matter and white matter.
Ⅳ. Discussion
Among the several techniques used to fill the k-space, the most commonly used technique is the cartesian trajectory. However, the cartesian technique generates various artifacts due to the involuntary motion of the patients in the phase encoding direction. In actual clinical setting, it is recommended to use the MultiVane technique instead of the cartesian technique to correct artifacts arising from the motion. Nevertheless, the MultiVane technique has not been commonly used because the time to create the images by filling the k-space becomes longer in the MultiVane technique. In this experiment, cartesian and multivane trajectories were compared and analyzed by using an ACR phantom in the absence of motion, although the MultiVane technique is for correcting artifacts arising from the motions in the cartesian trajectory. It would have been more helpful if we used a real phantom (Rando Phantom or anthropomorphic phantom) rather than an ACR phantom.
Ⅴ. Conclusions
According to the results, the geometric accuracy of both cartesian and MultiVane trajectories was within the normal range of 190±2 mm. The high-contrast spatial resolution in the cartesian trajectory was measured as 0.9 in both axial T1 and T2, which was within the normal range. The high-contrast spatial resolution in the MultiVane trajectory was measured as 1.0 in axial T1 and 0.9 for UL and 1.1 for RL in axial T2, in which RL was deviated from the normal range. The IIU was 0.15% higher in the cartesian trajectory than in the MultiVane trajectory in the axial T1. In axial T2, the cartesian trajectory showed 0.47% higher IIU than the MultiVane trajectory. The low-contrast object detectability was 38.4 spokes for the cartesian trajectory in axial T1, which falls under the normal range while low-contrast object detectability was 36.2 spokes for the MultiVane trajectory, which was deviated from the normal range. In axial T2, the value was 37.2 spokes for the cartesian trajectory, and 31.2 spokes for the MultiVane trajectory, in which the value in the cartesian trajectory falls under the normal range, while the value in the MultiVane trajectory was deviated from the normal range. As a result, the results of using the cartesian technique gave better results than the MultiVane technique in terms of the high-contrast spatial resolution, image intensity uniformity and low-contrast object detectability. This study would help obtain the basic data that can be used in actual clinical setting.